Operational Experience with the EM-3000™ Non-contact Specular Microscope

Copy Link
Published Online: Feb 16th 2011 US Ophthalmic Review, 2007,3:11-3 DOI: http://doi.org/10.17925/USOR.2007.03.00.11
Authors: Shigeru Kinoshita
Quick Links:
Article
Article Information
Article:

A non-contact specular microscope makes no direct contact with the eye and is effective in capturing images easily, two features that are important for clinical applications. However, there are some disadvantages, such as a narrower capturing range and lower image quality compared with a contact-type specular microscope. It has also been found that errors can arise from the automatic analysis of captured images. However, the Tomey EM-3000™ non-contact specular microscope mostly solves these problems. As we have had the opportunity to use the EM-3000 in our clinic, it is our pleasure to introduce this new instrument.

Features
The EM-3000 is an ‘all-in-one’ specular microscope that includes an in-built program to automatically analyze images of corneal endothelial cells, thus allowing physicians to easily perform automatic analyses without the use of a personal computer (see Figure 1).

The primary advantages of the EM-3000 include ease of operation and rapid image capture. The ‘one-touch’ panel of the EM-3000 allows for quick and easy operation of the instrument simply by touching the screen. The ‘autoalignment’ and ‘autoshot’ functions instantly capture images when the center of the pupil shown on the screen is touched as the capturing head moves toward the eye to be examined. In addition, since the capturing head moves automatically, the patient can remain stable even when switching the eye to be captured from right to left, thus the EM-3000 provides increased comfort for the patient. As the capturing time is only 1.2 seconds, the patient does not need to stop blinking. Moreover, since a series of 15 images are instantly captured, the number of capturing errors is small.

The light source of the EM-3000 is a long-life light-emitting diode (LED), not the conventional flashlight format, thus eliminating the need for any replacement lamps. This improvement enables continuous capturing and reduces out-of-focus images. When using a conventional non-contact specular microscope, a skilled physician is often needed to finely adjust the patient’s eye or face in order to obtain proper images. However, the EM-3000 allows images to be captured comparatively smoothly by anyone, thus freeing the skills of the physician for more important tasks.

An additional advantage of the EM-3000 is its capacity to analyze images soon after capture. The in-built automatic analysis software displays the cell density, cell area, flow co-efficient (CV) value, hexagonal cell appearance ratio, and corneal thickness in less than eight seconds. As the capturing range of 0.25×0.54mm is larger than that of conventional models, more cells can be analyzed. In addition, unique functions are also provided that allow the display mode to be changed for different purposes, with specific buttons such as ‘traced display’ to display the outline of the cell used for automatic analysis ‘area-specific color display’ to display cells in different colours that correspond to the area of each cell, and ‘shape-specific color display’ to display cells in different colors that correspond to the shape of each cell. Furthermore, auxiliary functions are also available for when automatic capturing or analysis is difficult for eyes with corneal diseases or immediately after corneal transplantation. If autoalignment is difficult, capturing can be performed in ‘manual’ mode. If there are many atypical cells in a captured image or part of the image is out of focus, the manual edit can edit the trace line or analyze when only the clear sections are available. The EM-3000 provides ease of operation, rapid capturing capability, and a large variety of normal and auxiliary analysis functions, and is useful for observing corneal endothelial cells.

Operational Experience for Clinical Applications
In December 2006 and July 2007, we used the EM-3000 to capture images of 112 eyes following penetrating keratoplasty (PKP) and 56 eyes with no keratoplasty—168 eyes in total (75 eyes of 41 male patients and 93 eyes of 59 female patients between 35 and 90 years of age, mean age 68±15 years)—at the Cornea Clinic of the Baptist Eye Clinic, Kyoto.

In post-keratoplasty eyes it is sometimes difficult to capture endothelial cell images even though the cornea is clear. However, in almost all cases it was easy to automatically capture images when using the EM-3000, and only a support was needed to keep the patient’s eye open. Automatic capturing was impossible in only a few cases, but the image could still be captured in manual mode. Historically, it has been difficult to capture a clear image of endothelial cells when using a conventional specular microscope due to factors such as an irregular posterior corneal surface; however, the improved design of the EM-3000 allows for images of those cells to be captured clearly, even in the days immediately following keratoplasty (see Figure 2).

Once the images are captured, the EM-3000’s built-in analysis software automatically calculates the cell density, cell area, and CV value. To reliably confirm the results of the automatic analysis, we assessed the calculated cell density calculated by manually tracing the shape of all endothelial cells visually recognized on the image that were captured as the true value of the cell density (manual value), and compared this with the density obtained by automatic analysis (automatic value) (see Figure 3).

Of the analyzable images taken by the EM-3000, we recognised a significantly high correlation for 132 of 168 eyes (see Figure 4). However, the difference between the manual value and automatic value was 500 pieces/mm2 or more for eight out of 132 eyes (7%) (see Figure 5). The corneal endothelial cell images of those eight eyes were all unclear, so the number of cells that were subject to automatic analysis was small (see Figure 6).

We could confirm that errors may occur in automatic analysis when the captured image is unsuccessful. In such cases, it is effective to delete unnecessary trace lines and/or to add correct trace lines using the edit function of the EM-3000.

Currently, we are in the process of trying to capture images of patients in the early stages after Descemet’s stripping automated endothelial keratoplasty (DSAEK) at the Kyoto Prefectural University Hospital. As the LED of the EM-3000 is superior to the conventional flashlight source for continuous capturing, it should allow us to capture good eye images. If so, the EM-3000 will be of further benefit to ophthalmology professionals.

Article Information:
Received

2011-02-16T00:00:00

Further Resources

Share this Article
Related Content In Imaging
  • Copied to clipboard!
    accredited arrow-down-editablearrow-downarrow_leftarrow-right-bluearrow-right-dark-bluearrow-right-greenarrow-right-greyarrow-right-orangearrow-right-whitearrow-right-bluearrow-up-orangeavatarcalendarchevron-down consultant-pathologist-nurseconsultant-pathologistcrosscrossdownloademailexclaimationfeedbackfiltergraph-arrowinterviewslinkmdt_iconmenumore_dots nurse-consultantpadlock patient-advocate-pathologistpatient-consultantpatientperson pharmacist-nurseplay_buttonplay-colour-tmcplay-colourAsset 1podcastprinter scenerysearch share single-doctor social_facebooksocial_googleplussocial_instagramsocial_linkedin_altsocial_linkedin_altsocial_pinterestlogo-twitter-glyph-32social_youtubeshape-star (1)tick-bluetick-orangetick-red tick-whiteticktimetranscriptup-arrowwebinar Sponsored Department Location NEW TMM Corporate Services Icons-07NEW TMM Corporate Services Icons-08NEW TMM Corporate Services Icons-09NEW TMM Corporate Services Icons-10NEW TMM Corporate Services Icons-11NEW TMM Corporate Services Icons-12Salary £ TMM-Corp-Site-Icons-01TMM-Corp-Site-Icons-02TMM-Corp-Site-Icons-03TMM-Corp-Site-Icons-04TMM-Corp-Site-Icons-05TMM-Corp-Site-Icons-06TMM-Corp-Site-Icons-07TMM-Corp-Site-Icons-08TMM-Corp-Site-Icons-09TMM-Corp-Site-Icons-10TMM-Corp-Site-Icons-11TMM-Corp-Site-Icons-12TMM-Corp-Site-Icons-13TMM-Corp-Site-Icons-14TMM-Corp-Site-Icons-15TMM-Corp-Site-Icons-16TMM-Corp-Site-Icons-17TMM-Corp-Site-Icons-18TMM-Corp-Site-Icons-19TMM-Corp-Site-Icons-20TMM-Corp-Site-Icons-21TMM-Corp-Site-Icons-22TMM-Corp-Site-Icons-23TMM-Corp-Site-Icons-24TMM-Corp-Site-Icons-25TMM-Corp-Site-Icons-26TMM-Corp-Site-Icons-27TMM-Corp-Site-Icons-28TMM-Corp-Site-Icons-29TMM-Corp-Site-Icons-30TMM-Corp-Site-Icons-31TMM-Corp-Site-Icons-32TMM-Corp-Site-Icons-33TMM-Corp-Site-Icons-34TMM-Corp-Site-Icons-35TMM-Corp-Site-Icons-36TMM-Corp-Site-Icons-37TMM-Corp-Site-Icons-38TMM-Corp-Site-Icons-39TMM-Corp-Site-Icons-40TMM-Corp-Site-Icons-41TMM-Corp-Site-Icons-42TMM-Corp-Site-Icons-43TMM-Corp-Site-Icons-44TMM-Corp-Site-Icons-45TMM-Corp-Site-Icons-46TMM-Corp-Site-Icons-47TMM-Corp-Site-Icons-48TMM-Corp-Site-Icons-49TMM-Corp-Site-Icons-50TMM-Corp-Site-Icons-51TMM-Corp-Site-Icons-52TMM-Corp-Site-Icons-53TMM-Corp-Site-Icons-54TMM-Corp-Site-Icons-55TMM-Corp-Site-Icons-56TMM-Corp-Site-Icons-57TMM-Corp-Site-Icons-58TMM-Corp-Site-Icons-59TMM-Corp-Site-Icons-60TMM-Corp-Site-Icons-61TMM-Corp-Site-Icons-62TMM-Corp-Site-Icons-63TMM-Corp-Site-Icons-64TMM-Corp-Site-Icons-65TMM-Corp-Site-Icons-66TMM-Corp-Site-Icons-67TMM-Corp-Site-Icons-68TMM-Corp-Site-Icons-69TMM-Corp-Site-Icons-70TMM-Corp-Site-Icons-71TMM-Corp-Site-Icons-72